Continuous-Flow Glucose Test Walkthrough

Continuous-Flow Glucose Test — Setup, Terminology, and Process Walkthrough

Three-reservoir, three-pump configuration · reconstructed from the rig schematic, 2026-11-18

This process changes the concentration inside the test vessel without disturbing it. The vessel stays in place with a constant fill volume while the solution is exchanged through its fixed ports, so the antenna sees the same vessel position, liquid level and geometry throughout the run. This minimizes perturbation to the test vessel and reduces or eliminates geometry as a confounding variable, so changes in the antenna response can be attributed to concentration.

1The setup

The rig holds a small test vessel at a constant fill volume while three peristaltic pumps move liquid through it, and a single collection container (R3) on a scale catches everything that leaves. R2 primes the vessel with prime solution (C_B0); R1 then dilutes it with diluent (C_A) while the displaced liquid is weighed. The vessel connects through two Y-connectors (tees) on its right side: Y1 on the lower port (shared by P1 diluent-in and P3 flush-out) and Y2 on the middle port (shared by P2 prime-in and the V2 dilution-outlet). A third top port is the vent / overflow through V1. Because the vessel never moves and stays full, the antenna sees the same geometry throughout, so changes in its response track concentration, not geometry.

R3 is the only collection point — there is no separate drain reservoir and no second container. Everything the vessel sheds (priming overflow, flush liquid and dilution effluent) lands in R3 on the scale. Tare the scale at the start of Dilution, so the mass counted toward concentration labels is the dilution effluent alone, not the earlier priming and flushing collected in the same container.

2Terminology

Containers and concentrations

SymbolMeaningRole
R1 / C_ADiluent source / its concentrationSupplies liquid during Dilution (pump P1)
R2 / C_B0Prime source / initial concentration in the vesselFills the vessel during Priming (pump P2)
R3Collection container, on the scaleReceives priming overflow, flush liquid and dilution effluent
Test vessel / V_BDilution chamber / its working volumeConstant working volume during dilution; where the antenna reads
C_B(w)Current concentration in the vesselA function of cumulative effluent volume w, not of time directly
C_BfThe vessel's concentration at the end of DilutionA snapshot of C_B(w) at the last sweep, not a separate container
wCollected dilution-effluent volumew = (scale mass since dilution start) ÷ density

Pumps and valves

SymbolConnectsRole
P1 (Dilute)R1 → Y1 (lower port)Diluent inflow during Diluting
P2 (Prime)R2 → Y2 (middle port)Prime inflow during Priming
P3 (Flush)Y1 (lower port) → R3Flush outflow during Flushing
V1 (Vent / overflow)top port → R3Open to vent / overflow during Priming & Flushing; closed to seal the vessel during Diluting
V2 (Dilution outlet)Y2 (middle port) → R3Open only during Diluting, to pass displaced effluent to R3
Y1 / Y2Y-connectors on the vessel's right sideY1 joins P1 & P3 at the lower port; Y2 joins P2 & V2 at the middle port

Open = valve open · Closed = valve closed. A pump is active only in the phase that drives it (P2 in Priming, P3 in Flushing, P1 in Diluting).

3The three processes, animated

PhaseActive pumpV1V2Flow to R3
PrimingP2OpenClosedOverflow from the top port via V1
DilutingP1ClosedOpenDisplaced liquid from Y2 (middle port) via V2
FlushingP3OpenClosedFlush liquid from Y1 (lower port) via P3
Phase

R1 C_A R2 C_B0 R3 · Collection 0.0 g scale · tare at dilution start collects overflow + flush + effluent Vessel V_B = 30 mL Y2 Y1 P1 · Dilute P2 · Prime P3 · Flush V1 vent / overflow V2 Vessel concentration Effluent volume w (mL) = scale reading / ρ C_B(w)
Effluent mass since tare, m
0.0 g
Effluent, w = m / ρ
0.0 mL
Volumes exchanged, w / V_B
0.00
Label, C_B
1000.0

4From scale weight to concentration label

The antenna sweeps continuously during Dilution. Each sweep is labeled with the vessel's concentration at that moment, worked out from the scale reading at the same timestamp:

  1. Tare the scale at the start of Dilution, so earlier priming overflow and flush liquid already in R3 are zeroed out.
  2. Read the net mass m (g) — the dilution effluent collected since the tare — at the sweep's timestamp.
  3. Convert to collected effluent volume: w = m / ρ (mL).
  4. Apply the dilution model: C_B(w) = C_A + (C_B0 - C_A) × exp(-w / V_B).
  5. Store that C_B as the sweep's label.

The model assumes the vessel is perfectly mixed and holds a constant volume V_B, with inflow (P1) equal to outflow (through V2). After one vessel volume of effluent (w = V_B), C_B has moved 63% of the way from C_B0 to C_A, and after about 4.6 volumes it is within 1%. Using w instead of elapsed time means a pump rate that drifts during the run doesn't change the labels.

Labeling table for the current inputs

Scale m (g)w (mL)w / V_BLabel C_B