This process changes the concentration inside the test vessel without disturbing it. The vessel stays in place with a constant fill volume while the solution is exchanged through its fixed ports, so the antenna sees the same vessel position, liquid level and geometry throughout the run. This minimizes perturbation to the test vessel and reduces or eliminates geometry as a confounding variable, so changes in the antenna response can be attributed to concentration.
The rig holds a small test vessel at a constant fill volume while three peristaltic pumps move liquid through it, and a single collection container (R3) on a scale catches everything that leaves. R2 primes the vessel with prime solution (C_B0); R1 then dilutes it with diluent (C_A) while the displaced liquid is weighed. The vessel connects through two Y-connectors (tees) on its right side: Y1 on the lower port (shared by P1 diluent-in and P3 flush-out) and Y2 on the middle port (shared by P2 prime-in and the V2 dilution-outlet). A third top port is the vent / overflow through V1. Because the vessel never moves and stays full, the antenna sees the same geometry throughout, so changes in its response track concentration, not geometry.
| Symbol | Meaning | Role |
|---|---|---|
| R1 / C_A | Diluent source / its concentration | Supplies liquid during Dilution (pump P1) |
| R2 / C_B0 | Prime source / initial concentration in the vessel | Fills the vessel during Priming (pump P2) |
| R3 | Collection container, on the scale | Receives priming overflow, flush liquid and dilution effluent |
| Test vessel / V_B | Dilution chamber / its working volume | Constant working volume during dilution; where the antenna reads |
| C_B(w) | Current concentration in the vessel | A function of cumulative effluent volume w, not of time directly |
| C_Bf | The vessel's concentration at the end of Dilution | A snapshot of C_B(w) at the last sweep, not a separate container |
| w | Collected dilution-effluent volume | w = (scale mass since dilution start) ÷ density |
| Symbol | Connects | Role |
|---|---|---|
| P1 (Dilute) | R1 → Y1 (lower port) | Diluent inflow during Diluting |
| P2 (Prime) | R2 → Y2 (middle port) | Prime inflow during Priming |
| P3 (Flush) | Y1 (lower port) → R3 | Flush outflow during Flushing |
| V1 (Vent / overflow) | top port → R3 | Open to vent / overflow during Priming & Flushing; closed to seal the vessel during Diluting |
| V2 (Dilution outlet) | Y2 (middle port) → R3 | Open only during Diluting, to pass displaced effluent to R3 |
| Y1 / Y2 | Y-connectors on the vessel's right side | Y1 joins P1 & P3 at the lower port; Y2 joins P2 & V2 at the middle port |
Open = valve open · Closed = valve closed. A pump is active only in the phase that drives it (P2 in Priming, P3 in Flushing, P1 in Diluting).
| Phase | Active pump | V1 | V2 | Flow to R3 |
|---|---|---|---|---|
| Priming | P2 | Open | Closed | Overflow from the top port via V1 |
| Diluting | P1 | Closed | Open | Displaced liquid from Y2 (middle port) via V2 |
| Flushing | P3 | Open | Closed | Flush liquid from Y1 (lower port) via P3 |
The antenna sweeps continuously during Dilution. Each sweep is labeled with the vessel's concentration at that moment, worked out from the scale reading at the same timestamp:
w = m / ρ (mL).C_B(w) = C_A + (C_B0 - C_A) × exp(-w / V_B).The model assumes the vessel is perfectly mixed and holds a constant volume V_B, with inflow (P1) equal to outflow (through V2). After one vessel volume of effluent (w = V_B), C_B has moved 63% of the way from C_B0 to C_A, and after about 4.6 volumes it is within 1%. Using w instead of elapsed time means a pump rate that drifts during the run doesn't change the labels.
| Scale m (g) | w (mL) | w / V_B | Label C_B |
|---|